Custom Silicone Diaphragms: Pressure Range, Thickness Control and Cycle-Life Testing

Custom silicone diaphragms are flexible components that separate two chambers while responding to changes in pressure, vacuum or mechanical force. They are used in pumps, valves, regulators, pressure sensors, respiratory equipment, dosing systems, household appliances and industrial fluid-control assemblies.

Unlike a static gasket, a diaphragm must repeatedly flex while maintaining a seal. Its performance depends on the interaction between material properties, effective diameter, thickness, profile, clamping geometry, pressure differential and operating stroke.

A diaphragm that is too thick may require excessive actuation force or respond too slowly. A diaphragm that is too thin may stretch, leak, rupture or fail prematurely under pressure cycling.

Reliable development therefore requires three connected controls:

  1. Defining the real operating and proof-pressure range
  2. Controlling thickness in every functional area
  3. Testing the completed assembly through representative pressure cycles

What Is a Silicone Diaphragm?

A silicone diaphragm is a thin, flexible membrane installed between two housing sections. Its outer edge is usually clamped or sealed while the central area moves in response to pressure.

Depending on the product, the diaphragm may perform one or several functions:

  • Separate air from liquid
  • Prevent contamination between chambers
  • Transfer pressure
  • Meter or pump fluid
  • Actuate a valve
  • Regulate pressure
  • Detect pressure changes
  • Provide a return force
  • Protect an electronic sensor
  • Compensate for volume changes
  • Isolate aggressive media from mechanical components

일반적인 용도로는 다음과 같은 것들이 있습니다:

  • Medical pumps
  • Ventilators
  • Anesthesia equipment
  • Respiratory devices
  • Infusion equipment
  • Laboratory analyzers
  • Water valves
  • 커피 머신
  • Soap dispensers
  • Pressure switches
  • Automotive fluid systems
  • Pneumatic controls
  • Chemical dosing pumps

Trelleborg identifies silicone and other elastomeric diaphragm seals as solutions for low-pressure valves and respiratory equipment. It also notes that fabric reinforcement can be used for higher-pressure applications. Trelleborg Medical Diaphragm Seals

Pressure Range Is Not a Single Material Value

A silicone grade does not have one universal diaphragm pressure rating.

The maximum safe pressure depends on:

  • Unsupported diameter
  • Diaphragm thickness
  • Diaphragm profile
  • 실리콘 탄성 계수
  • 인장 강도
  • 인열 강도
  • 연신율
  • Clamping method
  • Center support
  • Operating stroke
  • 온도
  • 유체 호환성
  • Pressure waveform
  • 수명 요구 사항
  • 안전 계수

The same LSR material may survive a relatively high pressure in a small, supported diaphragm but fail at a much lower pressure when used in a larger unsupported membrane.

For this reason, pressure range should be specified at the component and assembly level—not only on the silicone material datasheet.

Define the Pressure Conditions Correctly

Several pressure values should be established before diaphragm design begins.

Pressure termMeaning
Minimum operating pressureLowest pressure at which the diaphragm must respond correctly
Nominal operating pressureNormal working pressure during most of the product’s life
Maximum operating pressureHighest continuous or repeated working pressure
Differential pressurePressure difference between the two sides of the diaphragm
Proof pressureTemporary pressure the assembly must withstand without permanent damage
Burst pressurePressure at which rupture, pull-out or structural failure occurs
Vacuum conditionNegative pressure that may reverse the diaphragm or pull it against a housing surface
Pressure spikeShort transient pressure above normal operation
Pulsation rangeRepeated minimum-to-maximum pressure during cycling

A component designed for 100 kPa on one side and 90 kPa on the other experiences only 10 kPa differential pressure. A component exposed to 100 kPa on one side and vacuum on the other experiences a much larger differential.

The complete pressure history must therefore be considered.

Pressure Force and Effective Diameter

The approximate net force produced by a pressure differential can be estimated as:

Force = Differential pressure × Effective projected area

As effective diaphragm diameter increases, the area—and therefore the total pressure force—increases rapidly.

This means that a small increase in diameter can significantly increase:

  • Membrane stress
  • Center displacement
  • Clamping load
  • Required housing strength
  • Risk of pull-out
  • Burst-pressure demand

A larger diaphragm may provide greater sensitivity or displacement, but it generally requires more careful control of thickness and profile.

Main Silicone Diaphragm Configurations

Flat Diaphragm

A flat diaphragm is simple and economical. It can be suitable for limited movement and relatively small pressure differentials.

Potential limitations include:

  • High strain during large displacement
  • Strong sensitivity to clamping condition
  • Stress concentration near the fixed edge
  • Limited stroke before excessive stretching

Preformed or Dished Diaphragm

A shallow dome or dish provides additional material for movement.

Advantages can include:

  • Greater stroke
  • Reduced stretching
  • Lower actuation force
  • Better control of movement direction

The formed profile must remain stable during demolding, assembly and cycling.

Convoluted Diaphragm

One or more molded convolutions allow the diaphragm to flex through geometry rather than relying only on material stretching.

기대되는 이점은 다음과 같습니다:

  • Longer stroke
  • Lower spring rate
  • Reduced membrane strain
  • Improved cycle life

However, convolution depth, radii and wall thickness require precise control.

Rolling Diaphragm

A rolling diaphragm moves by rolling along a piston or housing surface.

It can provide:

  • Long, controlled stroke
  • Low friction
  • Low hysteresis
  • Good pressure response

The housing surface, alignment and diaphragm profile must be tightly controlled to prevent wrinkling or abrasion.

Fabric-Reinforced Diaphragm

A textile reinforcement layer increases dimensional stability and burst resistance.

Fabric-reinforced designs may be preferred for:

  • Higher pressure
  • Large diameters
  • Low stretch
  • Long stroke
  • Precise pressure response
  • Reduced ballooning

Trelleborg notes that higher pressure and burst resistance can be supported through synthetic-fabric reinforcement, while thinner homogeneous diaphragms can improve control sensitivity. Trelleborg Life Sciences Solutions

Insert-Molded Diaphragm

Metal or plastic inserts can be integrated into the diaphragm center or edge.

Insert molding may:

  • Simplify assembly
  • Improve force transfer
  • Provide a connection to a valve stem
  • Prevent center wear
  • Improve positional repeatability
  • Reduce secondary adhesive bonding

The insert should use chemical adhesion, mechanical locking or both.

How Thickness Affects Performance

Thickness is one of the most important diaphragm dimensions.

A small change can alter:

  • 구동력
  • Pressure sensitivity
  • Stroke
  • 히스테리시스
  • 응력 분포
  • Burst pressure
  • 응답 시간
  • Fatigue life

Thicker Diaphragms

Increasing thickness generally produces:

  • Higher stiffness
  • Lower displacement at the same pressure
  • Higher required actuation force
  • Improved resistance to pressure
  • Better dimensional stability
  • Lower risk of local tearing

잠재적인 단점으로는 다음과 같은 것들이 있습니다:

  • Reduced sensitivity
  • Shorter available stroke
  • Higher stress at transitions
  • Longer molding cure time
  • Difficult filling in adjacent thin areas

Thinner Diaphragms

Reducing thickness generally produces:

  • Lower actuation force
  • Greater pressure sensitivity
  • More displacement
  • Faster movement
  • Better response to small pressure differences

잠재적인 단점으로는 다음과 같은 것들이 있습니다:

  • Higher strain
  • Lower burst resistance
  • Increased ballooning
  • Greater sensitivity to thickness variation
  • Higher risk of pinholes
  • Greater risk of handling damage

In bending-dominated regions, stiffness can change approximately with the cube of thickness. However, a diaphragm often transitions into membrane stretching at larger deflections, so final behavior should be verified through simulation and physical testing.

Uniform Thickness vs. Designed Thickness Zones

Not every diaphragm should have one uniform thickness.

A functional design may include:

  • A thicker clamping bead
  • A thin flexible membrane
  • Reinforced transition zones
  • A thicker center pad
  • Flexible convolutions
  • A rigid insert interface

Thickness transitions should be gradual. Sudden changes can create local stress concentrations and become fatigue-initiation points.

The drawing should identify thickness requirements by functional zone rather than providing only one general thickness dimension.

Thickness-Control Challenges

Silicone diaphragms can be difficult to measure because the material is soft and flexible. Excessive probe force can compress the membrane and produce an incorrect result.

Possible measurement methods include:

  • Low-force thickness gauge
  • Non-contact optical measurement
  • Laser scanning
  • Optical comparator
  • Controlled sectioning of qualification samples
  • Coordinate measurement of supported parts
  • Mold-cavity verification
  • Part-weight monitoring as an indirect control

The inspection method should define:

  • 측정 위치
  • Support fixture
  • Contact force
  • Probe diameter
  • 컨디셔닝 시간
  • 온도
  • Number of points
  • Acceptance limits

Thickness variation near a convolution or radius may be more important than variation in a low-strain center area.

What Causes Thickness Variation?

Common manufacturing causes include:

  • Mold-cavity mismatch
  • 움직임 삽입
  • Uneven mold shut-off
  • Inconsistent material loading
  • 공기 포집
  • 이음선 플래시
  • Tool deflection
  • 불균일한 배기
  • 충전 불완전
  • Local cure variation
  • 금형 마모
  • Incorrect compression-molding charge position

For LSR injection molding, balanced gating, accurate cavity machining and stable process parameters are essential.

LSR Injection Molding vs. Compression Molding

LSR 사출 성형

Liquid silicone rubber injection molding is often selected for thin, complex or high-volume diaphragms.

장점은 다음과 같습니다:

  • Repeatable material metering
  • 자동화 생산
  • Thin-section filling
  • Complex profiles
  • 여러 개의 공동
  • 수작업 감소
  • Good suitability for cleanroom production
  • Integration with plastic or metal inserts

LSR processing uses controlled injection rate, pressure, mold temperature and cure time. Component geometry and material chemistry both influence cycle time. SIMTEC LSR Injection Molding Guide

압축 성형

Compression molding may be appropriate for:

  • HCR silicone
  • Lower production quantities
  • Larger diaphragms
  • Fabric-reinforced components
  • Development tooling
  • Selected thick or simple geometries

However, charge placement and compression flow can influence reinforcement position and thickness consistency.

Fabric-Reinforced Molding

Fabric must be accurately positioned before molding. Important controls include:

  • Fabric orientation
  • Tension
  • Preforming
  • Edge location
  • Silicone penetration
  • Wrinkle prevention
  • 삽입 위치
  • Cut-edge protection

A misaligned fabric layer can create asymmetric movement and inconsistent pressure response.

Silicone Material Selection

경도

A softer silicone may provide lower actuation force and greater movement. A harder material may improve dimensional stability and pressure resistance.

Hardness alone does not define diaphragm performance. Modulus at working strain is often more important than the Shore A value.

인장 강도 및 연신율

The material must tolerate repeated stretching without permanent damage.

ASTM D412 provides methods for evaluating tensile strength and elongation of vulcanized rubber, but standard test-sheet values do not replace actual diaphragm cycling. ASTM D412

인열 강도

Tear resistance is important near:

  • Clamping edges
  • 구멍
  • Insert interfaces
  • Gate-removal areas
  • Flash-trimmed edges
  • Sharp housing features

A very small notch can become a fatigue crack after thousands or millions of cycles.

압축 영구 변형

The sealing bead may remain compressed for the entire product life. Excessive compression set can reduce clamping force and cause leakage.

ASTM D395 evaluates the ability of rubber to retain elastic properties after prolonged compressive stress, although dynamic diaphragm life still requires application-specific testing. ASTM D395

Stress Relaxation

Even if the material does not show severe permanent deformation, clamping force may decrease over time. This can affect perimeter sealing and diaphragm retention.

Fluid Compatibility

The silicone must be tested with the actual operating media, including:

  • 스팀
  • Saline
  • Medication
  • Detergent
  • 석유
  • 연료
  • 세정제
  • Food ingredients
  • 소독제
  • Process gases

Exposure can cause swelling, softening, hardening, extraction or loss of mechanical strength.

온도

Temperature changes both the silicone properties and the fluid behavior.

The design should consider:

  • 최저 작동 온도
  • 최대 연속 사용 온도
  • Short temperature peaks
  • Sterilization temperature
  • 보관 온도
  • 열 사이클링

Regulatory Requirements

Depending on the application, requirements may include:

  • FDA food-contact compliance
  • USP Class VI
  • ISO 10993 evaluation
  • RoHS
  • REACH
  • Customer-specific extractables limits
  • Medical-device traceability
  • 클린룸 제조

The regulatory status of the raw material does not automatically certify the finished diaphragm for every application.

Diaphragm DFM Guidelines

Use Smooth Radii

Sharp internal corners concentrate stress. Use smooth radii where the flexible membrane joins:

  • Clamping beads
  • Center pads
  • Convolutions
  • 삽입물
  • 두꺼운 단면

Keep the Parting Line Away from Critical Flex Zones

Parting-line mismatch or flash can create a notch that grows during repeated movement.

Whenever possible, place the parting line away from the highest-strain region.

Control Gate Position

The gate should not create a weak point in the working membrane.

Avoid gate marks in areas exposed to:

  • Repeated bending
  • High stretch
  • Sealing contact
  • Fluid flow
  • Cosmetic inspection

Provide Effective Venting

Thin diaphragms can trap air at the end of fill. Poor venting may cause:

  • 단거리 샷
  • 핀홀
  • Burned areas
  • Weak knit lines
  • 두께 편차

Because LSR flows easily, vents must also be designed to limit flash.

Design the Clamping Bead Carefully

The perimeter bead may provide:

  • 밀봉
  • Positioning
  • Retention
  • Compression control

The housing should compress the bead sufficiently to seal without pinching the membrane or forcing it into the moving zone.

Avoid Excessive Assembly Stretch

A diaphragm should not be pulled significantly out of shape during installation unless controlled pre-tension is part of the design.

Uncontrolled stretch changes:

  • Effective thickness
  • Stroke
  • Pressure response
  • Fatigue life
  • 밀봉력

Housing Design and Assembly

The diaphragm and housing must be developed as one system.

Important housing factors include:

  • Clamp width
  • Clamp gap
  • 표면 마감
  • Edge radii
  • Alignment
  • Available stroke space
  • Vent paths
  • 패스너 간격
  • 평탄도
  • Assembly torque
  • Protection against overcompression

Uneven fastener torque can distort the sealing bead and change diaphragm movement.

If the diaphragm can contact a housing wall during operation, the contact surface should be smooth and free of sharp edges.

Pressure-Displacement Testing

A pressure-displacement test measures how far the diaphragm moves at different pressure levels.

The test can evaluate:

  • 민감도
  • Spring rate
  • 히스테리시스
  • Maximum stroke
  • Return position
  • 영구 변형
  • 구강 간 변동

A typical test records diaphragm displacement while pressure is increased and then decreased.

The loading and unloading curves may not be identical because silicone is viscoelastic. The difference between them is part of the diaphragm’s hysteresis.

Proof-Pressure Testing

Proof pressure confirms that the diaphragm can withstand a temporary pressure above normal operation without:

  • 누수
  • Rupture
  • Pull-out
  • 영구 변형
  • Insert separation
  • Loss of functional response

After proof testing, the component should be returned to normal conditions and tested again for displacement, leakage and appearance.

Burst-Pressure Testing

Burst testing determines the pressure at which structural failure occurs.

Possible failure modes include:

  • Membrane rupture
  • Tearing at the clamped edge
  • Insert pull-out
  • Bead extrusion
  • Housing separation
  • Fabric delamination
  • Leakage before rupture

Burst pressure should be tested using a protected fixture and controlled pressure ramp.

A production specification may require a minimum burst value rather than testing every part to destruction.

누출 시험

Leakage can occur through:

  • The silicone membrane
  • A pinhole
  • The perimeter seal
  • An insert interface
  • A molded knit line
  • Housing deformation

Common leak-test methods include:

  • Pressure decay
  • Vacuum decay
  • 기포 시험
  • Direct flow measurement
  • Tracer-gas testing
  • Liquid hold testing

The method should define:

  • Test media
  • 압력
  • 안정화 시간
  • 시험 소요 시간
  • 온도
  • Allowable leakage
  • Fixture volume
  • Pass/fail calculation

Cycle-Life Testing

Cycle-life testing is essential because many diaphragm failures develop gradually.

A diaphragm may pass initial proof-pressure and leakage tests but fail after repeated flexing.

Define a Realistic Pressure Waveform

The cycle should reproduce the actual application as closely as possible.

Possible waveforms include:

  • Minimum-to-maximum pressure
  • Positive pressure to vacuum
  • Pulsed pressure
  • Constant displacement
  • Mechanical actuation
  • Pressure with fluid flow
  • Pressure dwell followed by rapid release

The test should not automatically use a simple maximum-to-zero cycle if the real product behaves differently.

Define Cycle Frequency

Running a test too quickly can generate heat and create a failure mechanism that would not occur in normal service.

Cycle frequency should consider:

  • Actual operating speed
  • Silicone heat buildup
  • Fluid heating
  • Valve response time
  • Test-fixture limitations

Accelerated testing is useful only when the acceleration method does not change the fundamental failure mode.

Define the Mission Profile

A complete cycle requirement should state:

  • Total number of cycles
  • 압력 범위
  • Displacement or stroke
  • Frequency
  • 작동 주기
  • 체류 시간
  • 온도
  • Test media
  • 조립 상태
  • Inspection intervals

Measure Performance During Cycling

Do not wait until the final cycle to inspect the diaphragm.

Periodic checks may include:

  • 누수
  • Pressure-displacement response
  • Return position
  • Visual cracking
  • 영구 변형
  • 움직임 삽입
  • Thickness change
  • 구동력

Define the Failure Criteria

Possible failure criteria include:

  • Visible crack
  • Pin-hole leakage
  • Excessive change in displacement
  • Reduced pressure response
  • 영구 변형
  • Insert separation
  • Perimeter leakage
  • Failure to return
  • Burst-pressure reduction

Factors That Reduce Cycle Life

Excessive Strain

Large displacement in a flat diaphragm can produce high membrane strain. A convolution or preformed shape may reduce strain.

Sharp Transitions

Small radii and abrupt thickness changes create localized stress.

Surface Defects

Flash, trimming damage, scratches and contamination can initiate cracks.

Incorrect Clamping

Overcompression or edge pinching can damage the flexible transition.

Material Aging

Heat, chemicals, sterilization and UV exposure may change silicone properties.

마모

Contact with a rough housing surface can gradually wear through the diaphragm.

Insert Stress

A rigid center insert can create high strain around its edge if the transition is not properly designed.

Pressure Spikes

Short pressure events may exceed the intended stress level even when average operating pressure is low.

Simulation and Finite Element Analysis

Finite element analysis can help compare diaphragm designs before tooling.

Useful outputs include:

  • 응력 분포
  • Strain concentration
  • Displacement
  • Contact with the housing
  • Clamping effects
  • 로딩 중
  • Pressure response

The material model should represent the hyperelastic behavior of the selected silicone.

Possible models include:

  • Mooney-Rivlin
  • Ogden
  • Neo-Hookean
  • Yeoh

Accurate simulation requires material test data over the expected strain range. A generic elastic modulus is usually insufficient for large diaphragm deformation.

Simulation should guide prototype design, but physical testing remains essential.

Typical Validation Sequence

Stage 1: Material Screening

Compare candidate silicone grades for:

  • 경도
  • 모듈러스
  • 인장 강도
  • 연신율
  • 인열 강도
  • 압축 영구 변형
  • 유체 호환성
  • 규제 현황

Stage 2: Prototype Geometry

Evaluate controlled variations in:

  • 막 두께
  • Effective diameter
  • Convolution depth
  • Center-pad size
  • Clamping bead
  • 도형 삽입

Stage 3: Functional Testing

측정:

  • Pressure-displacement response
  • 히스테리시스
  • 누수
  • Proof pressure
  • Burst pressure
  • 구동력

Stage 4: Environmental Conditioning

Retest after:

  • 열 노화
  • 유체 침지
  • 열 사이클링
  • 멸균
  • 저장 숙성
  • 습도에의 노출

Stage 5: Cycle-Life Testing

Run the production-intent diaphragm inside the production-intent housing.

6단계: 시범 생산

확인:

  • Thickness consistency
  • Cavity-to-cavity performance
  • 삽입 위치
  • Assembly repeatability
  • 검사 능력
  • Test-fixture stability
  • 추적성

일반적인 고장 유형 및 시정 조치

문제가능한 원인시정 조치
Diaphragm ruptures at centerMembrane too thin or pressure too highIncrease thickness or add support
Tearing occurs at outer edgeSharp clamping edge or excessive strainIncrease radius and revise clamp design
Pressure response is too lowMembrane too thick or material too hardReduce stiffness or change profile
Diaphragm moves too farMembrane too flexible or effective area too largeIncrease thickness or add a travel stop
Perimeter leakageInsufficient bead compression or uneven housingRevise bead, clamp gap and assembly torque
Cycle life is inconsistentThickness or surface-defect variationImprove tooling and cavity-level inspection
Diaphragm does not returnMaterial set, excessive stretch or vacuum lockReview material, profile and venting
Insert separatesPoor bonding or high edge stressAdd mechanical locking and improve transition
Fabric wrinklesPoor preforming or locationImprove fabric fixture and molding process
Cracks form after sterilizationMaterial degradation or residual stressValidate the material and post-cure process
Diaphragm sticks to housingSurface contact or fluid residueIncrease clearance or modify surface finish

Production Quality Control

A diaphragm production-control plan may include:

  • Raw-material lot traceability
  • LSR mixing-ratio monitoring
  • Mold-temperature records
  • Cure-time control
  • Cavity identification
  • Part weight
  • Critical thickness measurements
  • Membrane-profile inspection
  • 플래시 검사
  • Insert-position inspection
  • Perimeter-bead dimensions
  • Pressure-displacement sampling
  • 누출 검사
  • Proof-pressure sampling
  • Periodic cycle-life verification
  • Cleanliness and packaging inspection

For multi-cavity tools, thickness and functional data should be reviewed by cavity.

RFQ Checklist for Custom Silicone Diaphragms

Provide the following information for an accurate quotation and DFM review:

  • 2D 도면
  • 3D CAD 파일
  • Diaphragm type
  • Effective diameter
  • Available installation space
  • Required stroke
  • Minimum operating pressure
  • Nominal operating pressure
  • Maximum operating pressure
  • Vacuum condition
  • Proof pressure
  • Burst-pressure requirement
  • Pressure waveform
  • Cycle-life target
  • Cycle frequency
  • Fluid or gas type
  • 작동 온도
  • Sterilization or cleaning method
  • Silicone hardness preference
  • 규제 요건
  • Reinforcement requirements
  • 재료 삽입
  • Required leakage rate
  • Thickness tolerance
  • Housing drawing
  • 조립 방법
  • 시제품 수량
  • 연간 수요
  • Inspection-report requirements
  • 포장 및 청결 기준

Avoid providing only a nominal pressure value. The manufacturer also needs the effective area, maximum differential pressure, stroke, temperature, media and cycle requirement.

자주 묻는 질문

What pressure can a silicone diaphragm withstand?

There is no universal pressure rating. Pressure capability depends on effective diameter, thickness, profile, material, clamping method, temperature and required cycle life.

Does increasing thickness always improve diaphragm life?

Not always. Greater thickness can improve pressure resistance but may increase stiffness and stress at transition areas. The correct thickness balances movement, strain and pressure.

Is LSR suitable for very thin diaphragms?

Yes. LSR can fill thin and complex geometries, but tooling precision, venting, flash control and thickness inspection become increasingly important.

When is fabric reinforcement required?

Fabric reinforcement may be considered for larger diameters, higher pressure, longer strokes, reduced ballooning or more stable pressure response.

Can a diaphragm be molded with a metal insert?

Yes. Metal or plastic inserts can be overmolded into the center or perimeter. Bonding and mechanical locking should be validated through pull, leak and cycle testing.

What is the difference between proof pressure and burst pressure?

Proof pressure confirms that the part survives a specified overload without damage. Burst pressure identifies the pressure at which structural failure occurs.

How many cycles should a diaphragm be tested?

The requirement depends on the product mission profile. The test should represent expected service life plus an appropriate validation margin.

Can cycle-life testing be accelerated?

Yes, but excessive test frequency, temperature or pressure can introduce unrealistic failure modes. The acceleration method should be technically justified.

Why must the diaphragm be tested inside the housing?

The housing controls clamping, compression, clearance and available stroke. A loose diaphragm may perform differently from the assembled component.

How is thickness measured on a soft diaphragm?

Low-force contact gauges or non-contact optical methods are preferred. The measurement locations, support fixture and contact force must be defined.

결론

Custom silicone diaphragm performance depends on more than selecting a silicone hardness and nominal thickness.

Pressure rating, displacement, sealing and cycle life are controlled by the complete system:

  • Effective diameter
  • Membrane profile
  • Thickness distribution
  • Silicone material
  • Edge radii
  • Clamping bead
  • Housing geometry
  • Operating pressure
  • 온도
  • Media exposure
  • Cycle waveform

The most reliable development process combines DFM, material testing, nonlinear simulation, prototype molding, pressure-displacement measurement, proof and burst testing, environmental conditioning and full cycle-life validation.

For a custom silicone diaphragm project, provide the operating pressure range, diaphragm diameter, required stroke, media, temperature, housing design and cycle-life target. These details allow the manufacturer to recommend the appropriate LSR grade, thickness profile, reinforcement method, tooling process and validation plan.